Closed-Cup Flash Point Tester

Product origin CHINA

Delivery time Within 7 days

Supply capacity Factory Direct Supply

Application:
It is used to make determination of the closed cup flash point of the petroleum products.

  • Closed-Cup Flash Point Tester

    Request a Call Back

    •  Ask a Question

      Closed-Cup Flash Point Tester

      Ask a Question


        Share
      Closed-Cup Flash Point Tester
      Features: BK-FP261: * The heating power is adjustable and shown visually by voltmeter. * The structure is designed small and exquisite with stainless-steel table board. BK-FP261D: * 8-inch IPS high-definition capacitive screen, the test data and parameters are displayed by curve in real time. * Equipped with Bluetooth interface. The user can query test data anytime. * Temperature rise, cover opening, ignition, detection and printing data are completed automatically, and the test arm rises and falls automatically. * Remote automatic upgrade to obtain the latest version, remote prediction, abnormal early warning, comprehensive evaluation and maintenance of instrument operation status. * Electronic ignition, gas flame, automatic air cooling at the end of the test. Technical Parameters:
      Model BK-FP261 BK-FP261D
      Ambient Temperature 15~35°C
      Relative Humidity ≤85%
      Flash Point Determination Mercury-in-glass thermometer. Specifications are as below: Scale -5°C~110°C, division 0.5°C Scale 20°C~150°C, division 1°C Scale 90°C~370°C, division 2°C Range: Ambient to 230°C Repeatability:≤3°C Reproducibility: ≤4°C Accuracy: 0.1 °C
      Oil Cup Inner diameter: 50.7mm~50.8mm. Depth: 55.7mm~56.0mm The scribed line depth of capacity of testing oil: 33.9mm〜34.3mm. Capacity of testing oikabout 70ml
      Stirring Rate Procedure A: (90~120)RPM, Procedure B: (250±10)RPM
      Igniting Device Igniting source: gas(or other civilian fuels, the same below) Electric ignition. Diameter of igniter: 0.7mm~0.8mm
      Heating Device The heating power is continuously adjustable from 0 Wto 600W.
      Heating Rate (1~12)°C/min: Controllable and adjustable Procedure A: (5-6)°C/min, Procedure B: (1~1.5)°C/min, Procedure C:(3±0.5) °C/min
      Consumption M650W ≤600W
      Power Supply AC220V±10%, 50Hz
      Dimension(W*D*H) 370*320*300mm 510*320*330mm
      Package Size(W*D*H) 650*450*360mm 690*500*595mm
      Gross Weight 19kg 34kg