      Features: pH: 1. 1 to 5 points calibration with automatic recognition for USA, NIST and DIN buffers 2. Automatic electrode diagnosis shows the pH slope and offset 3. Solution temperature coefficient compensates for the pure water samples and references the pH to 25°C ORP: 1. 1 point offset calibration allows adjusting the displayed value to a known standard 2. Relative and absolute millivolt modes ensure the reliable ORP measurements Ion Concentration: 1. 2 to 5 points calibration, including the 8 concentration points can be selected 2. Electrode management is capable of storing and recalling up to 3 electrode slopes 3. Automatic electrode diagnosis shows the calibration points and slopes 4. Selectable ion measurement methods (direct reading, known addition, known subtraction, sample addition, sample subtraction) and concentration units (ppm, mg/L, mol/L or mmol/L) Water Hardness: 1. 2 to 5 points calibration from low to high concentrations 2. Selectable measurement units (German degree, English degree, French degree, mmol/L and mg/L) are used for professional water hardness measurements General Features: 1. Automatic temperature compensation ensures accurate readings over the entire range 2. Auto-Read function senses and locks the measurement endpoint 3. Timed Interval Readings sends the measurement data to a PC or printer 4. Limit Alarm automatically alerts when reading exceeds the specified range 5. Calibration due alarm prompts user to calibrate the meter regularly 6. Password protection prevents the unauthorized calibration and settings 7. Expanded memory stores or recalls up to 1000 data sets 8. Reset function automatically resumes all settings back to the factory defaults Technical Parameters: 
      Model PH-A120-CN PH-A120-UK PH-A130-CN PH-A130-UK PH-A131
      PH Range -2.000~20.000pH
      Resolution 0.001, 0.01, selectable
      Accuracy ±0.002pH
      Calibration Points 1 to 5 points
      pH Buffer Options USA, NIST, DIN, 5 custom buffers
      ORP(optional) mV Range ±2000.0mV
      Relative mV Range ±2000.0mV
      Resolution 0.1mV
      Accuracy ±0.2mV
      Calibration Points 1 point
      Ion Range 0.001~30000 (depending on the range of ISE)
      Resolution 0.001, 0.01, 0.1, 1
      Accuracy ±0.5% F.S. (monovalent), ±1% F.S. (divalent)
      Measurement Units ppm, mg/L, mol/L, mmol/L
      Calibration Points 2 to 5 points (0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000)
      Measurement Methods Direct reading, known addition, known subtraction, sample addition, sample subtraction
      Electrode Management 1 to 3 electrodes
      Water Hardness Range 0.05~200mmol/L, 0~1122°dH, 0~1404°e, 0~2000°fH, 0~8000mg/L (Ca2+)
      Resolution 0.001, 0.01, 0.1, 1
      Accuracy ±1% F.S.
      Calibration Points 2 to 5 points (0.01, 0.1, 1, 10, 100mmol/L)
      Temperature Range 0~105°C/32~221°F
      Resolution 0.1°C/0.1°F
      Accuracy ±0.5°C/±0.9°F
      Offset Calibration 1 point, reading ±10°C
      Other Specifications Temperature Compensation 0~100°C, manual or automatic
      Solution Temperature Coefficient 25°C
      Stability Criteria Standard or high-accuracy
      Calibration Due Alarm 1 to 31 days or off
      Interval Readings 10, 30, 60 seconds, 10, 30 minutes or off
      Password Protection 4 digits
      Memory 1000 data sets
      Communication Interface USB
      Connector BNC, 3.5 mm jack socket
      Display 7 inches TFT LCD
      Power Supply Standard: 220V 50/60Hz; Optional: 110V, 50/60Hz;  DC12V power adapter
      Outer Dimensions 240*220*80mm
      Net Weight 1.7kg
      Packing Size 320*250*200mm
      Gross Weight 1.8kg
      Remarks: • means with; — means without. Difference of PH-A120/130-CN& PH-A120/130-UK: 1. PH-A120/130-CN electrode is E201-BNC plastic body pH electrode, PH-A120/130-UK electrode is P11 glass pH electrode 2. PH-A120/130-UK is more accurate than PH-A120/130-CN Ordering Information: 1. PH-A120/130-CN: Meter, E201-BNC plastic body pH electrode, temperature probe, pH buffer sachets, electrode holder and power adapter 2. PH-A120/130-UK: Meter, P11 glass pH electrode, temperature probe, pH buffer sachets, electrode holder and power adapter 3. PH-A131: Meter, ion selective electrode, temperature probe, standard solutions (100/1000ppm), ionic strength adjuster, electrode holder and power adapter