Auto ELISA Processor KAZKAZ2001

Product origin CHINA

Delivery time Within 7 days

Supply capacity Factory Direct Supply

Model 2001

Introduction:
Auto ELISA Processor KAZKAZ2001 is a fully automaticdetection equipment that integrates sample addition,incubation, shaking, washing, and reading. lt is widely usedin hospital laboratories, disease prevention and controlcenters, blood centers, and blood banks, animal and plantinspection and quarantine institutions, scientific researchinstitutions, university laboratories and other places.

  • Auto ELISA Processor KAZKAZ2001

    Request a Call Back

    •  Ask a Question

      Auto ELISA Processor KAZKAZ2001

      Ask a Question


        Share
      Features: 1. Adopt the principle of pressure type liquid detection 2. With TIP detection, clot detection function 3. High-precision pipetting 4. Position sensing, real-time monitoring of gripper status 5. Multi-spacing and multi-strength adjustable grippers can be adapted to grasp and place microplates and plate covers of different sizes 6. Dual thread control, can run simultaneously with the sampling arm without interfering with each other elisa Technical Parameters:
      Model BIOBASE2001
      Sample&Reagent Unit Sample Racks 6*20 push-pull sample rack with detection
      Sample Channel 2 channels
      Liquid Level Detection Air pressure level detection, clot detection function
      Tip 1000μl, 3 tip racks
      Pipetting 10μl~1000μl
      Pipetting Precision 100μl: CV≤1%, Accuracy≤2.5% 1000μl: CV≤0.5%, Accuracy≤1%
      Reagent Position 4 reagent racks, reagent tanks with detection
      Washing Unit Washer 1 unit
      Washing Head 16-channel(32-pin), double row cleaning, hanging installation, with leakage recovery
      Injection Uniformity CV<1.5%
      Washing Residual <1μl
      Reading Unit Reading Channels 8 independent photometric channels
      Linear Range 0.000~3.000Abs
      Repeatability CV≤1.0%
      Stability ≤±0.003Abs
      Optical Filters 4 standard filters(405, 450, 492, 630nm); Total 10 filters positions
      Reading Accuracy ≤±0.02Abs from 0.0 to 1.0OD ≤±0.03Abs from 1.0 to 2.0OD
      Sensitivity ≥0.01mg/L
      Channel Difference ≤0.02Abs
      Incubation Unit Incubators 8 shaking incubators
      Cover Positions 1(8 covers)
      Heating Method Dry heating with cover
      Shaking Function Horizontal circular oscillation, 1-3 levels adjustable, amplitude 2.5mm
      Temperature Range RT to 60 °C
      Temperature Uniformity 0.5℃
      Temperature Fluctuation ±0.5℃
      Data Management Software Above Windows 7 system
      LIS System Bi-direction, support HL7 protocol
      Results Absorbance and results reviewable by software
      Printer External optional
      Built-in Scanner Optional
      Other Function Biosafety Protection Optional (Negative pressure working zone, HEPA filter)
      Disinfection UV lamp
      Lighting LED
      IAP Function Firmware can be upgraded online at any time
      Working Condition Power Supply AC220V±10%, 50/60Hz(Standard); 110V±10%, 60Hz( Optional)
      Temperature 10~40℃
      Humidity 30~80%
      Size&Weight External Size (W*D*H) 1275*750*1650mm
      Net Weight 330kg
      Package Size(W*D*H)mm Machine: 1445*1130*920 Base Stand: 1445*1005*920 Accessories: 1065*510*895
      Gross Weight(kg) Machine: 220 Base Stand: 125 Accessories: 30